Operation and Maintenance Manual
WA R N I N G
Ozone is a hazardous gas and a strong oxidant. Exposure to ozone should be minimized
by using the instrument in a well-ventilated area and by venting the exhaust of the vac-
uum pump to a fume hood. The ozone generator should be turned off when the instrument
is not in use.
WA R N I N G
Burner temperature Is extremely hot. Do not touch. Allow to cool before servicing.
WA R N I N G
Hydrogen is an extremely flammable gas. Use appropriate care when handling. Inspect
all connections with a suitable leak detector.
WA R N I N G
Oxygen rich environments can promote combustion and even result in spontaneous com-
bustion under conditions of high pressure and exposure to contamination. Use only oxy-
gen rated components and ensure that components are oxygen clean prior to use with
pure oxygen.
WA R N I N G
Exceeding the gas inlet pressure of 25 psig (1.72 bar) may damage the hydrogen and oxi-
dant sensors or burst their connective tubing.