Operation and Maintenance Manual

Figure 13 RV5 Oil- Sealed Vacuum Pump and Associated Traps (Top)

Pump exhaust

Inlet & valve

Gas ballast

Pump inlet

Oil fill caps

Retractable handle

Chemical

Trap

N O T E

Sometimes water condensation and accumulation are visible in the exhaust line. This is
normal. However, do not allow water to continue to accumulate after approximately one
week of operation. Significant water accumulation may indicate improper pump operation,
an improperly vented exhaust line, or a Burner leak. Water accumulation in the exhaust line
can cause damage to the pump, especially if allowed to fall back into the pump. Contact
Agilent for advice if water accumulation continues to occur.