4-11
IM MW180-01E
Calibrator
1
2
3
4
5
App
Index
• 6-Channel, Medium-Speed, Four-Wire RTD Resistance Input Module (MX110-V4R-M06)
Measurement Range to Be Calibrated
Input Value 1
Input Value 2
20 mV
0 mV on Ch1
20 mV on Ch2
60 mV
0 mV on Ch1
60 mV on Ch2
200 mV
0 mV on Ch1
200 mV on Ch2
1 V
0 V on Ch1
1 V on Ch2
2 V
0 V on Ch1
2 V on Ch2
6 V
0 V on Ch1
6 V on Ch2
20 V
0 V on Ch1
20 V on Ch2
100 V
0 V on Ch1
100 V on Ch2
RTD(1 mA)20 mV
0 Ω on Ch3
20 Ω on Ch4
RTD(1 mA)60 mV
0 Ω on Ch3
60 Ω on Ch5
RTD(1 mA)200 mV
0 Ω on Ch3
200 Ω on Ch6
RTD(1 mA)600 mV
0 Ω on Ch3
300 Ω on Ch4
RTD(0.25 mA)600 mV
0 Ω on Ch3
2400 Ω on Ch5
RTD(0.25 mA)1 V
0 Ω on Ch3
3000 Ω on Ch6
• 4-CH, Medium-Speed Strain Input Module (MX112-B12-M04, MX112-B35-M04,
MX112-NDI-M04)
Measurement Range to Be Calibrated
Connection Value 1
Connection Value 2
2000 µSTR (strain)
120.000 Ω on Ch2
117.154 Ω on Ch2
20000 µSTR (strain)
120.000 Ω on Ch2
113.010 Ω on Ch2
200000 µSTR (strain)
120.000 Ω on Ch2
80.000 Ω on Ch2
Use a 4-gauge method connection when performing calibration. For information on
this connection, see section 2.4 of the MW100 Data Acquisition Unit User’s Manual (IM
MW100-01E).
• 8-CH, Medium-Speed Analog Output Module (MX120-VAO-M08)
Adjust so that all channels output 0 V and 10 V. This differs from other input modules.
Range to Be Calibrated
Output Value 1
Output Value 2
10 V
0 V
10 V
Note
• For a description of the measurement range and accuracy of each input module, see
chapter 5 in the MW100 Data Acquisition Unit User’s Manual (IM MW100-01E).
• For procedures and wiring for calibration, see the MW100 Data Acquisition Unit User’s
Manual (IM MW100-01E), section 4.4.
4.2 Calibration Procedure