Explanation, Calibrated parameters – Yokogawa PC-Based MX100 User Manual
Page 59
![background image](/manuals/360294/59/background.png)
4-10
IM MW180-01E
4.2 Calibration Procedure
Explanation
Calibrated Parameters
• 4-CH, High-Speed Universal Input Module (MX110-UNV-H04)
Measurement Range to Be Calibrated
Input Value 1
Input Value 2
20 mV
0 mV
20 mV
60 mV
0 mV
60 mV
200 mV
0 mV
200 mV
1 V
0 V
1 V
2 V
0 V
2 V
6 V
0 V
6 V
20 V
0 V
20 V
100 V
0 V
100 V
RTD(1 mA)200 mV
0 Ω
200 Ω
RTD(1 mA)600 mV
0 Ω
300 Ω
RTD(1 mA)1 V
0 Ω
500 Ω
RTD(2 mA)60 mV
0 Ω
30 Ω
RTD(2 mA)200 mV
0 Ω
100 Ω
RTD(2 mA)600 mV
0 Ω
300 Ω
RTD(2 mA)1 V
0 Ω
250 Ω
• 10-CH, Medium-Speed Universal Input Module (MX110-UNV-M10)
Measurement Range to Be Calibrated
Input Value 1
Input Value 2
20 mV
0 mV on Ch1
20 mV on Ch2
60 mV
0 mV on Ch1
60 mV on Ch2
200 mV
0 mV on Ch1
200 mV on Ch2
1 V
0 V on Ch1
1 V on Ch2
2 V
0 V on Ch1
2 V on Ch2
6 V
0 V on Ch1
6 V on Ch2
20 V
0 V on Ch1
20 V on Ch2
100 V
0 V on Ch1
100 V on Ch2
RTD(1 mA)60 mV
0 Ω on Ch3
60 Ω on Ch6
RTD(1 mA)200 mV
0 Ω on Ch3
200 Ω on Ch4
RTD(1 mA)600 mV
0 Ω on Ch3
300 Ω on Ch5
• 30-CH, Medium-Speed DCV/TC/DI Input Module (MX110-VTD-L30)
Measurement Range to Be Calibrated
Input Value 1
Input Value 2
20 mV
0 mV on Ch1
20 mV on Ch2
60 mV
0 mV on Ch1
60 mV on Ch2
200 mV
0 mV on Ch1
200 mV on Ch2
1 V
0 V on Ch1
1 V on Ch2
2 V
0 V on Ch1
2 V on Ch2
6 V
0 V on Ch1
6 V on Ch2
20 V
0 V on Ch1
20 V on Ch2
100 V
0 V on Ch1
100 V on Ch2