Yokogawa Data Acquisition with PID Control CX2000 User Manual
Page 61
8-9
IM 04L31A01-01E
Procedures for Disassembly (CX1000)
8
8.
Remove the four screws from the inside corners of the LCD as shown.
9.
Remove the two contact segments one at a time by grasping them with
tweezers.
Lump Cable
Bracket
Connector
FFC
Contact segment
Contact segment
10.
Remove the FFC and lump cable from their connectors.
11.
Slowly lift the bracket upward to remove the bracket assembly.
Note
•
Take care not to damage any components on the inverter or signal processing boards.
•
Be careful not to damage the FFC.
LCD assembly with the bracket assembly removed
Contact segment
12.
Being careful not to damage any surrounding parts, grasp the backlight and
slowly pull it out.
Note
•
If the backlight is broken, be careful not to touch any glass shards when removing it. Also,
check for glass shards that may have fallen into the lamp holder or other areas, and
remove them.
•
Dirt and grime on the backlight and LCDs can cause unevenness in display intensity. Be
sure to keep these devices clean as you handle them.
LCD assembly with the backlight removed
8.4 Replacing the Backlight