Yokogawa Data Acquisition with PID Control CX2000 User Manual
Page 25
3-5
SM 04L31A01-01E
Adjustments
3
Calibration Using Communications
The calibration process proceeds as follows:
1.
Enter Calibration mode
2.
Perform calibration
3.
Store the calibrated values
4.
Return to operation mode
The commands necessary for calibration are listed below.
For the input connection method, see "Input Range Calibration" on page 3-2.
DSp1 Switch to setting mode
p1
0
Switch to operation mode
1
Switch to Set up mode
2
Switch to A/D calibration mode
XZp1,p2,p3
Perform A/D calibration (Cal/Exe)
p1
Group
For control input: 01—05 for the CX1000, and 01—05 or 06—10 for the
CX2000
For measurement input: 01—06 for the CX1000, and 01—10 or 11—20
for the CX2000
p2
Type of operation
Cal/Exec
p3
Calibration range
20 mV, 60 mV, 200 mV, 1 V, 2 V, 6 V, 20 V, Pt100
XZp1,p2,p3,p4,p5
Manual correction of A/D calibration value (Display)
p1
Group
For control input: 01—05 for the CX1000, and 01—05 or 06—10 for the
CX2000
For measurement input: 01—06 for the CX1000, and 01—10 or 11—20
for the CX2000
p2
Type of operation
Display
p3
Calibration range
20 mV, 60 mV, 200 mV, 1 V, 2 V, 6 V, 20 V, Pt100
p4
Zero input value
p5
Full scale input value
XZp1,p2,p3
Store the calibrated values (exit)
p1
Group
For control input: 01—05 for the CX1000, and 01—05 or 06—10 for the
CX2000
For measurement input: 01—06 for the CX1000, and 01—10 or 11—20
for the CX2000
p2
Type of operation: End
p3
Data save selection
STORE Save to a file
ABORT Don’t save
XZp1,DISPLAY?
Get information for calibrated values
p1
Group
For control input: 01—05 for the CX1000, and 01—05 or 06—10 for the
CX2000
For measurement input: 01—06 for the CX1000, and 01—10 or 11—20
for the CX2000
3.1 Calibration of the Measuring Instrument's Input