Ascii output format, see, 12 app – Yokogawa Touch Screen GP20 User Manual
Page 95
2-81
IM 04L51B01-17EN
Commands and Responses
1
2
App
2.10.31 Instrument’s Unit Configuration Information (_UNS or _UNR)
The output in response to the command “_UNS” or “_UNR” is shown below.
Syntax
EA
p1,p2,p3,p4,p5,p6,p7,p8,p9,p10
p1,p2,p3,p4,p5,p6,p7,p8,p9,p10
…
EN
Output Example
EA
Main,0,'GX20-1J',123456789,xx-xx-xx-xx-xx-xx,R1.01,/MT /C2,0,10,0
EN
One line (p1 to p10) contains configuration information of a single unit.
p
n
Value
Description
p1 Main, Sub
Unit dependency (main or sub) information. Fixed to “Main.”
p2 0,1
Unit address number. Fixed at 0.
p3 ’GX20-1J’,
’COMM-LAN’
Unit name (model name). Enclosed in single quotation marks.
P4 123456789
Product serial number.
p5 xx-xx-xx-
xx-xx-xx
MAC address.
p6 R1.01
Firmware version. The output format is “R+version.”
p7 /MT /C2
Options. Codes of installed options delimited by spaces.
p8 0
Fixed at 0.
p9 10
Maximum number of installable modules. If there are not installable
modules, 0 is output.
p10 0
Unit status. The unit status is output in a character string.
2.10 ASCII Output Format