beautypg.com

Yokogawa AXFA14G/C User Manual

Page 96

background image

IM 01E20C02-01E

9-4

9. ACTUAL OPERATION

Touch the

switch twelve times to move the

cursor to “Ext Auto Zero (I)”.

Touch the

switch to select “Ext Auto Zero

(Zero adjustment via external status input)”.

In order to request confirmation, the entire display

flashes on and off. Touch the

switch once

again at this time to fix selection of the automatic
zero adjustment function.

NOTE

When no operations are carried out for 20
seconds in the flashing state, the system will
automatically return to the Sub-item Parameter
Search Mode.

When the operations except

are carried

out, the parameter cannot be set.

The system automatically returns to sub-item
selection screen (E).

Zero adjustment will be started if the DIO terminals are shorted (when the active mode has
been set to “Closed/Short Act” using F23: DIO Active Mode).

This process will end after approximately 30 seconds.

Auto zero adjustment function is being executed.
(about 30 seconds.)

NOTE

When the DIO terminals continue to be shorted, this zero adjustment is automati-
cally repeated.

F0902-2.EPS

F:Status Function

20:DO Function

21:DIO Function

22:DO Active Mode

F21:DIO Function

Ext Auto Zero (I)

Sub-item

Parameter

Search Mode

F21:DIO Function
No Function

Ext Auto Zero (I)

Ext Ttl Preset (I)

Major item

Sub-item

selection (E)

Parameter

Replacement

(execute) Mode

F21:DIO Function
No Function

No Function

Alarm Output

X12

F0902-3.EPS

Now Auto Zero

Executing...