Yokogawa AXFA14G/C User Manual
Page 39
IM 01E20C02-01E
6-10
6. PARAMETER DESCRIPTION
T0610-2.EPS
Item
Name
Display unit
(BRAIN)
Data range
Display unit
/BRAIN
Default value
(*): Indicated item
Units
R/W
Position
of decimal
point
Description
Power Synch
(POWER SYNCH)
No
Yes
W
2
Memo 3
(MEMO 3)
R
Memo 1
(MEMO 1)
Power Frequency
(POWER FREQ)
Memo 2
(MEMO 2)
W
47.00 to 63.00
Hz
ASCII 16 characters
Yes
ASCII 16 characters
W
ASCII 16 characters
R/W
—
W
Software Rev No
(SOFTWARE REV)
Good
Error
R
J50
J60 —
(SELF CHECK)
See “6.5 Alarm Functions”.
Memo field
Memo field
Memo field
Software revision number
J42
Selects whether or not the internal
frequency is to be synchronized
with the power supply frequency.
50.00
J41
J30
J31
J40
Displays the power-supply
frequency (for Power Synch =
“Yes”), or sets the power-supply
frequency (for Power Synch=“No”).
(10) Item K (Menu K): Diagnostic Function Setting items
Menu K contains items that are relevant to the diagnosis of insulation adhesion to the electrode.
T0611.EPS
Item
Name
Display unit
(BRAIN)
Data range
Display unit
/BRAIN
Default value
(*): Indicated item
Units
R/W
Position
of decimal
point
Description
2
2
2
Adhesion Level3
(ADH LEVEL3)
Adhesion Level2
(ADH LEVEL2)
M ohm
Adhesion Level4
(ADH LEVEL4)
Adhesion Level1
(ADH LEVEL1)
Adh Measure Value
(ADH MEAS VAL)
Adhesion Check
(ADHESION CHK)
Diagnosis
(DIAGNOSIS)
0.00 to 100.00
0.00 to 100.00
M ohm
M ohm
0.00 to 100.00
M ohm
W
W
W
0.00 to 100.00
No
Yes
W
W
2
Good
Error
R
—
(SELF CHECK)
K15
K60
K14
Sets the resistance value for adhesion
Level 4 to the electorode.
Sets the resistance value for adhesion
Level 2 to the electorode.
Sets the resistance value for adhesion
Level 1 to the electorode.
Sets the resistance value for adhesion
Level 3 to the electorode.
R
K13
Displays the resistance value for
adhesion to the electrode.
See “6.5 Alarm Functions”.
1.00
K12
0.10
0.50
2
—
M ohm
3.00
K00
K10
K11
Selects whether or not to perform diagnosis
of adhesion to the electrode.
No
(11) Item M (Menu M): Automatic Zero-Adjustment Function Setting items
Menu M contains items that are relevant to automatic zero adjustment.
Item
Name
Display unit
(BRAIN)
Data range
Display unit
/BRAIN
Default value
(*): Indicated item
Units
R/W
Position
of decimal
point
Description
T0612.EPS
Auto Zero Exe
(AUTOZERO EXE)
R/W
Magflow Zero
(MAGFLOW
3
-99.999 to 99.999
Adjustment
(ADJUSTMENT)
0.000
No Execution
Execution
W
Good
Error
R
—
(SELF CHECK)
Displays the result of the automatic
zero adjustment, or sets the zero point.
See “6.5 Alarm Functions”.
No Execution
Selects whether or not automatic
zero adjustment is carried out.
Linked with B50.
M10
M11
M00
M60