11-1

11. OUTLINE

STANDARD SPECIFICATIONS

Refer to IM 01E20F02-01E for F

OUNDATION

Fieldbus

communication type and IM 01E20F12-01E for
PROFIBUS PA communication type marked with “

Converter

*1: Select two points from: one pulse output, one alarm

output, one status input, or two status outputs.

*2: For models without an indicator, the hand-held

terminal is necessary to set parameters.

Excitation Method: (Combined with AXF Remote
Flowtube)

• Standard dual frequency excitation:

Size 2.5 to 400 mm (0.1 to 16 in.)

• Enhanced dual frequency excitation:

Size 25 to 200 mm (1.0 to 8.0 in.)
(Optional code HF1 or HF2)

Input Signal (*1) “

”:

One Status Input: Dry contact
Load resistance: 200

Ω or less (ON), 100 kΩ or more

(OFF)

Output Signals “

”:

• One Current Output: 4 to 20 mA DC (load resistance:

750

Ω maximum, including cable resistance)

• One Pulse Output (*1):

Transistor contact output (open collector)
Contact capacity: 30 V DC (OFF), 200 mA (ON)
Output rate: 0.0001 to 10,000 pps (pulse/second)

• One Alarm Output (*1):

Transistor contact output (open collector)
Contact capacity: 30 V DC (OFF), 200 mA (ON)

• Two Status Outputs (*1):

Transistor contact output (open collector)
Contact capacity: 30 V DC (OFF), 200 mA (ON)

Communication Signals “

”:

BRAIN or HART communication signal
(Superimposed on the 4 to 20 mA DC signal)
Distance from power line: 15 cm (6 in.) or more (Parallel
wiring should be avoided.)

BRAIN:

Communication Distance:

Up to 1.5 km (0.93 miles), when polyethylene insulated
PVC-sheathed cables (CEV cables) are used.
Communication distance varies depending on the type
of cable and wiring used.

Load Resistance:

250 to 450

Ω (including cable resistance)

Load Capacitance: 0.22

µF or less

Load Inductance: 3.3 mH or less

Input Impedance of Communicating Device:

10 k

Ω or more (at 2.4 kHz)

11. OUTLINE

HART:

Load Resistance:

250 to 600

Ω (including cable resistance)

Note: HART is a registered trademark of the HART

Communication Foundation.

Data Security During Power Failure:

Data (parameters, totalizer value, etc.) storage by
EEPROM. No back-up battery required.

Indicator (*2):

Full dot-matrix LCD (32

ϫ132 pixels)

Lightning Protector “

”:

The lightning protector is built into the excitation current
output, the current output, the signal common, pulse/
alarm/status input and output terminals. When optional
code A is selected, the lightning protector is built into the
power terminals.

Protection:

General-purpose Use/Sanitary Type/TIIS Flameproof
type: IP66, IP67
Explosion proof type except TIIS:

In case of explosion proof type except TIIS, refer to
description of “Enclosure” in “HAZARDOUS AREA
CLASSIFICATION”.

Coating:

Case and Cover: Corrosion-resistant coating

Coating Color; Mint green (Munsell 5.6 BG 3.3/2.9 or its

equivalent)

Converter Material:

Case and Cover: Aluminum alloy

Mounting/Shapes:

• Mounting: 2-inch pipe
• Electrical Connection: ANSI 1/2 NPT female

ISO M20

ϫ1.5 female

JIS G1/2 (PF1/2) female

• Terminal Connection: M4 size screw terminal

Grounding:

Grounding resistance 100

Ω or less

When optional code A is selected, grounding resistance
10

Ω or less shall be applied.

* In case of explosion proof type except TIIS, follow the

domestic electrical requirements as regulated in each
country.

* In case of TIIS Flameproof type, refer to description of

“HAZARDOUS AREA CLASSIFICATION”.

Combined Remote Flowtube:

• AXFA14 Converter can be combined with size 2.5 to 400

mm (0.1 to 16 in.) of AXF Remote Flowtube.

• If a combined converter is changed from AXFA14 to

AXFA11 or vice versa, a new meter factor must be
adjusted by flow calibrations.

• In case that size 250 mm (10 in.) or larger is used in low

conductivity or high concentration slurries, please use the
AXFA11 Converter.