Zero Adjustment Using Display Unit Switches
This section describes the procedure for zero adjustment using the display unit switches.
(For more details regarding setting methods using these switches, refer to Chapter 5: Basic
Operating Procedures.)
The parameters for zero adjustment are B50/M10: Auto Zero Exe (and either of these
can be used to carry out this procedure). For more details regarding these parameters, refer
to Chapter 6: Parameter Description.
The parameter M10: Auto Zero Exe will be used in the following description.
Once in Setting Mode, use the
switch to
move the cursor to M: Adjustment.
Touch the
switch to access Sub-item
Parameter Search Mode.
Upon selection of M: Adjustment, the cursor will
be positioned at M10: Auto Zero Exe. (Sub-item
selection (D))
Touch the
switch to access Parameter
Replacement Mode.
Touch the
switch to move the cursor to
“Execution”.
Touch the
switch to select the “Execution”.
In order to request confirmation, the entire display
flashes on and off. Touch the
switch once
again at this time to fix selection of the automatic
zero adjustment function.
NOTE
When no operations are carried out for 20
seconds in the flashing state, the system will
automatically return to the Sub-item Parameter
Search Mode.
When the operations except
are carried
out, the parameter cannot be set.
Setting Mode
Major item
parameter
Setting Mode
K:Diagnosis
M:Adjustment
N:Test
M10:Auto Zero Exe
No Execution
No Execution
Execution
M10:Auto Zero Exe
No Execution
Execution
No Execution
Major item
Sub-item
selection (D)
M:Adjustment
10:Auto Zero Exe
11:Magflow Zero
M10:Auto Zero Exe
Execution
Sub-item
Parameter
Search Mode
Parameter
Replacement
(execute) Mode
Start:
Major Item
Parameter
Search Mode