4 iecex, Iecex -4 – Yokogawa AXFA14G/C User Manual

Page 115

background image

IM 01E20C02-01E

12-4

12. EXPLOSION PROTECTED TYPE INSTRUMENT

(3) Operation

For CSA C22.2 Series

WARNING

WARNING : OPEN CIRCUIT BEFORE REMOV-

ING COVER.
OUVRIR LE CIRCUIT AVANT
D’ENLEVER LE COUVERCLE.

• Take care not to generate mechanical spark

when access to the instrument and peripheral
devices in hazardous location.

For CSA E79 Series

WARNING

WARNING : AFTER DE-ENERGIZING, DELAY

20 MINUTES BEFORE OPENING.
APRÉS POWER-OFF, ATTENDRE
20 MINUTES AVANT D’OUVRIR.

• Take care not to generate mechanical spark

when access to the instrument and peripheral
devices in hazardous locations.

(4) Maintenance and Repair

WARNING

The instrument modification or parts replacement
by other than authorized representative of
YOKOGAWA Electric Corporation or
YOKOGAWA Corporation of AMERICA is
prohibited and will void Canadian Standards
Explosionproof Certification.

Conduit

Magnetic Flowmeter

Sealing Fitting

HAZARDOUS LOCATIONS

50 cm Max.

F1203.EPS

Figure 12.3.1 Conduit Wiring

12.4 IECEx

WARNING

Only trained persons use this instrument in
industrial locations.

(1) Technical Data

Applicable Standard:

IEC60079-0: 2004, IEC60079-1: 2003,
IEC61241-0: 2004, IEC61241-1: 2004,
IEC60529: 1999 + Edition 2.1: 2001

Certificate: IECEx KEM 05.0018

IECEx Flameproof Type

Ex d IIC T6
Maximum power supply voltage: 250 Vac/130 Vdc
Excitation Circuit: 140V max
Enclosure: IP66, IP67
Ambient Temp.: –40

°C to +60°C (–40°F to +140°F)

IECEx Type of Protection “Dust”

Ex tD A21 IP6x T95

°C

Maximum power supply voltage: 250 Vac/130 Vdc
Excitation Circuit: 140V max
Enclosure: IP66, IP67
Maximum surface temperature: T95

°C (+203°F)

Ambient Temp.: –40

°C to +60°C (–40°F to +140°F)

(2) Installation

WARNING

• All wiring shall comply with local installation

requirements and local electrical code.

• In hazadous locations, the cable entry devices

shall be of a certified IECEx flameproof type,
suitable for the conditions of use and correctly
installed.

• Unused apertures shall be closed with suitable

flameproof certified blanking elements. (The
plug attached is certified as the flameproof and
IP66 or IP67 as a part of this apparatus.)

• In case of ANSI 1/2 NPT plug, ANSI hexagonal

wrench should be applied to screw in.