12. EXPLOSION PROTECTED TYPE INSTRUMENT

(3) Operation

WARNING

• “OPEN CIRCUIT BEFORE REMOVING COV-

ERS.”

• “SEAL ALL CONDUITS WITHIN 18 INCHES”

in hazardous locations.

• When installed in Division 2, “SEALS NOT

REQUIRED”

• Take care not to generate mechanical spark

when access to the instrument and peripheral
devices in hazardous locations.

(4) Maintenance and Repair

WARNING

The instrument modification or parts replacement
by other than authorized representative of
Yokogawa Electric Corporation is prohibited and
will void the approval of Factory Mutual Re-
search Corporation.

Conduit

Magnetic Flowmeter

Sealing Fitting

HAZARDOUS LOCATIONS

18" (457 mm) Max.

F1202.EPS

Figure 12.2.1 Conduit Wiring

12.3 CSA

(1) Technical Data

Applicable Standard:

For CSA C22.2 Series;

C22.2 No 0, C22.2 No 0.4, C22.2 No 0.5,
C22.2 No 25, C22.2 No 30, C22.2 No 94,
C22.2 No 1010.1

For CSA E79 Series;

CAN/CSA-E79-0, CAN/CSA-E79-1,

Certificate: 1481213

For CSA C22.2 Series

Explosion proof for Class I, Division 1, Groups A,

B, C & D.

Dust-ignition proof for Class II/III, Division 1,

Groups E, F & G.

“SEAL ALL CONDUITS WITHIN 50 cm OF THE

ENCLOSURE”

“WHEN INSTALLED IN DIV. 2, SEALS NOT

REQUIRED”

Maximum power supply voltage: 250 Vac/130 Vdc
Excitation Circuit: 140V max
Enclosure: Type 4X
Temperature Code: T6
Ambient Temp.: –40

°C to +60°C (–40°F to +140°F)

For CSA E79 Series

Flameproof for Zone 1, Ex d IIC T6

Maximum power supply voltage: 250 Vac/130 Vdc
Excitation Circuit: 140V max
Enclosure: IP66, IP67
Temperature Code: T6
Ambient Temp.: –40

°C to +60°C (–40°F to +140°F)

(2) Installation

For CSA C22.2 Series

WARNING

• All wiring shall comply with Canadian Electrical

Code Part I and Local Electrical Codes.

• In hazardous location, wiring shall be in conduit

as shown in Figure 12.3.1.

WARNING : SEAL ALL CONDUITS WITHIN

50cm OF THE ENCLOSURE’.
UN SCELLEMENT DOIT ÉTRE
INSTALLÉ À MOINS DE 50cm DU
BOÎTIER.

• When installed in Division 2, “SEALS NOT

REQUIRED”

For CSA E79 Series

WARNING

• All wiring shall comply with local installation

requirements and local electrical code.

• In hazadous locations, the cable entry devices

shall be of a certified flameproof type, suitable
for the conditions of use and correctly installed.

• Unused apertures shall be closed with suitable

flameproof certified blanking elements. (The
plug attached is flameproof certified.)