Life Threatening Rhythm Alarms

When LTA monitoring is enabled, the Propaq MD will monitor for the following life
threatening ECG rhythms: asystole, ventricular fibrillation, ventricular tachycardia, extreme
bradycardia, and extreme tachycardia.

Depending upon configuration of the unit, it generates visible and audible alarms as needed.

Note:

LTA monitoring is not available in AED mode.

Equipment Alert Display

When a problem with the Propaq MD unit or an attached sensor triggers an alert, in addition to
sounding an equipment alert, the Propaq MD unit displays an alert message (yellow
background, black text).

Warning!

Always respond immediately to a system alarm since the patient may not be monitored
during certain alert conditions.

In the following example, an equipment alert message indicates that the SpO

2

sensor has

become unattached (

SpO2 Check Sensor

) from the unit: