CPR Dashboard

9650-0806-01 Rev. C

Propaq MD Operator’s Guide

19–3

Constant Metronome Disabled

AED Mode

If Enable Metronome is set to Always or AED Mode Only, the metronome is active when
chest compressions are first detected during the CPR interval.

If set to Never, the metronome is silent.

Advisory/CPR Protocol or Manual Mode

If Enable Metronome is set to Always, the metronome is active only when the reported
compression rate falls below 80 cpm during the CPR period. If the chest compression sensor
is connected, the metronome is suspended after compressions have not been detected for at
least 2 seconds.

If set to Never or AED Mode Only, the metronome is silent.

The following table indicates when the CPR metronome is active.

CPR Dashboard

Whenever CPR-D-padz or OneStep electrodes

are connected to the Propaq MD defibrillator

and it senses compression, the unit illuminates the CPR Dashboard in the lower center portion
of the display. This field includes the CPR feedback indicators described in the following
sections.

Enable
Metronome
Setting

AED Mode

Advisory/CPR
Protocol Mode

Manual Mode

Constant Metronome Enabled

AED Mode
Only

On when CPR detected
Off when CPR Idle

Off

Off

Always

On when CPR detected
Off when CPR Idle

On when CPR detected
Off when CPR Idle

On when CPR detected
Off when CPR Idle

Never

Off

Off

Off

Constant Metronome Disabled

AED Mode
Only

On at start of CPR period

Off

Off

Always

On at start of CPR period

On only at <80
compressions per minute

On only at <80
compressions per minute

Never

Off

Off

Off