Turn on unit, 1 select energy level – ZOLL Propaq MD Rev C User Manual
Page 193

Emergency Defibrillation Procedure with Hands-Free Therapy Electrodes
9650-0806-01 Rev. C
Propaq MD Operator’s Guide
15–7
Note:
If it is not possible to place the “BACK” electrode on the patient’s back, place the
electrodes in the standard apex-sternum positions. Effective defibrillation results, but
pacing will usually be less effective.
Turn On Unit
Press the green power switch located on the top of the unit. The green, yellow, and red lights at
the top of the unit flash on and off, and the unit displays the message SELF TEST PASSED.
If defibrillation electrodes are not making good contact with the patient’s skin and the lead
selection is ECG Lead, the unit issues the message CHECK THERAPY ELECTRODES and
does not allow the delivery of energy.
1 Select Energy Level
Press the Select Energy arrows up or down to select the desired energy level. These buttons are
located on the front panel of the unit.
The following are the default energy selections:
•
Adult - 120 joules
•
Pediatric- 50 joules
•
Neonate - 50 joules
Note:
Neonatal and pediatric defibrillator energy levels should be selected based on site-
specific protocols.
The selected energy level is shown on the display.
Pad
Skin