Common tasks, Navigation keys, Display brightness – ZOLL Propaq MD Rev C User Manual
Page 57

Common Tasks
9650-0806-01 Rev. C
Propaq MD Operator’s Guide
2–17
Navigation Keys
Use the navigation keys (up/clockwise arrow, down/counterclockwise arrow, and select button)
to navigate through windows and make selections.
Using Up/Clockwise and Down/Counterclockwise Arrows
Use the up/clockwise down/counterclockwise arrows to do the following:
•
Move clockwise and counterclockwise through the main display windows.
•
Move up and down in a window.
•
Change parameter settings.
Using the Select Button
Use the Select button to do the following:
•
Display the settings window while a parameter is highlighted in the main window.
•
Select options from a window.
Display Brightness
The monitor can display in two different brightness modes:
•
high contrast with white background (for optimal display in bright sunlight)
•
color with black background (numerics and waveforms are easy to read)
•
night vision goggle (NVG) friendly (display and alarm LEDs prevent interference with
goggles)
Common Tasks
The section contains procedures for the following tasks:
•
“Changing the Display Brightness” on page 2-18.
•
“Replacing a Battery Pack on the Propaq MD” on page 2-18.
•
“Using Treatment Buttons” on page 2-19
Disarm
Safely discharges the defibrillator internally. No energy is delivered to
the patient.
IBP Setup
Brings up the IBP Control Panel for the corresponding channel (P1,
P2, or P3).
IBP Zero
Zeroes the IBP transducer for the corresponding channel (P1, P2, or
P3).
Table 2-3.
Propaq MD Quick Access Keys
Quick access key
Description
Disarm
Disarm
Disarm