NIBP System Messages
When monitoring NIBP, the Propaq MD unit may display the following messages:
System Message
Cause
READING IN PROGRESS
The unit is taking an NIBP measurement and
functioning normally.
READING STOPPED
The unit has stopped an NIBP measurement,
because the operator has pressed the NIBP button
and cancelled the measurement.
READING FAILED
The patient’s pulse is too weak to obtain an NIBP
measurement, or the cuff requires adjustment.
ARTIFACT
Excessive artifact is preventing the NIBP
measurement. If ECG is available, consider turning
SmartCuf on.
NEONATE ALERT
The unit has detected a neonate cuff when in Adult
mode. Replace cuff or correct patient mode, as
appropriate. If the detection is in error, dismiss the
alert and reattempt the NIBP measurement.
AIR LEAK
A major air leak is preventing cuff inflation. Check
hose and cuff connections, replace a defective hose
or cuff, as necessary, and reattempt NIBP
measurement.
KINKED HOSE
An air obstruction is preventing the correct operation
of the unit. Check hose for kinks. If using a dual
lumen hose, verify the correct placement of the O
ring on the hose connector. Eliminate kinks or
replace hose, if necessary.
NIBP DISABLED
A system error has occurred, and the Propaq MD
unit should be serviced.
PLEASE WAIT
After completing an NIBP request in Automatic mode
or STAT (TurboCuf) mode, the unit must wait up to
30 seconds before initiating the next NIBP
measurement request.
PRESS. LIMIT EXCEEDED
The cuff pressure has exceeded the allowable limit.
Restrict patient movement and check the hose for
kinks. Eliminate kinks or replace the hose, if
necessary.