beautypg.com

Pulse oximeter – ZOLL Propaq MD Rev C User Manual

Page 304

background image

Specifications

A-20

www.zoll.com

9650-0806-01 Rev. C

Pulse Oximeter

Range:

Oxygen Saturation (% SpO

2

)

1% - 100%

Carboxyhemoglobin Saturation (% SpCO)

0% - 99%

Methemoglobin Saturation (% SpMet)

0% - 99%

Pulse Rate (bpm)

25 - 240 beats per minute

Accuracy:

Oxygen Saturation (% SpO

2

) - During No Motion Conditions

Adults, Pediatrics

1

70% - 100%, ±2 digits 0% - 69%, unspecified

Neonates

2

70% - 100%, ±3 digits 0% - 69%, unspecified

Oxygen Saturation (% SpO

2

) - During Motion Conditions

3

Adults, Pediatrics

70% - 100%, ±3 digits 0% - 69%, unspecified

Neonates

70% - 100%, ±3 digits 0% - 69%, unspecified

Oxygen Saturation (% SpO

2

) - During Low Perfusion Conditions

4

Adults, Pediatrics

70% - 100%, ±2 digits

Neonates

70% - 100%, ±3 digits

Carboxyhemoglobin Saturation (% SpCO)

5

1% - 40% ±3 digits

Methemoglobin Saturation (% SpMet)

5

1% - 15% ±1 digit

Pulse Rate (bpm) - During No Motion Conditions

1

Adults, Pediatrics, Neonates

25 - 240 ±3 digits

Pulse Rate (bpm) - During Motion Conditions

3

Adults, Pediatrics, Neonates

25 - 240 ±5 digits

Resolution: SpO

2

: 1%

SpCO: 1%
SpMet: 0.1% for range up to 9.9%, 1% for range 10 - 99%
Pulse rate: 1 bpm (beats per minute)

Alarm Limits: On/Off displayed on monitor. User selectable.

SpO

2

: High 72 - 100% saturation, Low 70 - 98% saturation

SpCO: High 2 - 100% saturation, Low 0 - 99% saturation
SpMet: High 1 - 100% saturation, Low 0 - 99% saturation
Pulse Rate: High 60 - 235 beats per minute, Low 20 - 100 beats per minute

SpO

2

Wavelength for LNCS Sensors:

Nominal Red LED Wavelength: 660 nanometers
Nominal Infrared LED Wavelength: 905 nanometers

Energies (Radiant Power) of light for
LNCS Sensors at 50 mA pulsed:

15 mW

SpO

2

Wavelength for Rainbow Sensors:

The Rainbow sensors use 8 different LEDs with wavelengths of 610 - 905
nanometers