ZOLL Propaq MD Rev C User Manual
Page 313

Electromagnetic Compatibility Guidance and Manufacturer’s Declaration
9650-0806-01 Rev. C
Propaq MD Operator’s Guide
A-29
Electromagnetic Compatibility Guidance and
Manufacturer’s Declaration
In-Flight Use (RTCA/DO-160):
The Propaq MD unit complies with RTCA/DO-160, Environmental Conditions and Test
Procedures for Airborne Equipment, using the methods in Section 21, Category M for Radiated
and Conducted Radio Frequency Energy.
Guidance and manufacturer’s declaration – electromagnetic emissions
The Propaq MD unit is intended for use in the electromagnetic environment specified below. The customer or
the user of the Propaq MD unit should assure that it is used in such an environment.
Emissions test
Compliance
Electromagnetic environment – guidance
RF Emissions CISPR 11
Group 1
The Propaq MD unit uses RF energy only for its internal
function. Therefore, its RF emissions are very low and are
not likely to cause any interference in nearby electronic
equipment.
RF emissions CSPR 11
Class B
Harmonic emission IEC 6100-
3-2
Class A
The Propaq MD unit is suitable for use in all
establishments, including domestic establishments and
those directly connected to the public low-voltage power
supply network that supplies buildings used for domestic
purposes.
Voltage fluctuations/ flicker
emissions IEC 61000-3-3
Complies
Medical Electrical Equipment needs special precautions regarding EMC and needs to be installed and put
into service according to EMC information provided in this document.