Electromagnetic Immunity (IEC 60601-1-2)

Guidance and manufacturer’s declaration – electromagnetic immunity

The Propaq MD unit is intended for use in the electromagnetic environment specified below. The customer
or the user of the Propaq MD should assure that it is used in such an environment.

Immunity test

IEC 60601 test level

Compliance level

Electromagnetic environment
– guidance

Electrostatic
discharge (ESD) IEC
61000-4-2

± 6 kV contact
± 8 kV air

± 6 kV contact
± 8 kV air

Floors should be wood, concrete
or ceramic tile. If floors are
covered with synthetic material,
the relative humidity should be at
least 30%.

Electrical fast
transient / burst
IEC 61000-4-4

± 2 kV for power
supply lines

± 1 kV for input/
output lines

± 2 kV for power
supply lines

± 1 kV for input/
output lines

Mains power quality should be
that of a typical commercial or
hospital environment.

Surge
IEC 61000-4-5

± 1 kV line(s) to line(s)

± 2 kV line(s) to earth

± 1 kV differential
mode
± 2 kV common mode

Mains power quality should be
that of a typical commercial or
hospital environment.

Voltage dips, short
interruptions and
voltage variations on
power supply input
lines
IEC 61000-4-11

<5% U

T

(>95% dip in U

T

) for

0.5 cycle

40% U

T

(60% dip in U

T

) for

5 cycles

70% U

T

(30% dip in U

T

) for

25 cycles

<5% U

T

(>95% dip in U

T

) for

5 sec

<5% U

T

(>95% dip in U

T

) for

0.5 cycle

40% U

T

(60% dip in U

T

) for

5 cycles

70% U

T

(30% dip in U

T

) for

25 cycles

<5% U

T

(>95% dip in U

T

) for

5 sec

Mains power quality should be
that of a typical commercial or
hospital environment. If the user
of the Propaq MD unit requires
continued operation during
power mains interruptions, it is
recommended that the Propaq
MD unit be powered from an
uninterruptible power supply or a
battery.

Power frequency
(50/60 Hz) magnetic
field
IEC 61000-4-8

3 A/m

3 A/m

Power frequency magnetic fields
should be at levels characteristic
of a typical location in a typical
commercial or hospital
environment.

Note: U

T

is the AC mains voltage prior to application of the test level.