Troubleshooting the Detector

NOTE: Make sure the wiring from the detector to the Scintillator Input Remote I/O
Module is correct. If you have just replaced (or added) a Scintillator Input Remote I/O
Module when the Detector Failure occurred, refer to the Remote I/O Module Addressing
section to ensure you have properly addressed the Remote I/O Module.

• Remove the cable from the detector. This is a military style twist-off connector

located at the top of the detector.

• Make sure the Remote I/O Modules are powered up; the red LED on the Scintillator

Remote I/O Modules needs to be On or Flashing.

• Check the voltage at pins A & J of the twist-off connector. If you have greater than

14VDC at these pins, the detector needs to be repaired or replaced. Otherwise,
replace or repair the detector cable.