Replace Test Source

To initially install (or eventually replace) a Test Source, you use this menu. Typically,
the information required here is listed on a tag on the Test Source. This information is
necessary for the system to determine the amount of source activity decay between
Detector Sensitivity Checks to determine a proper Allowance Percent for the check.

• Click on the arrow (on the top box) and select the type (isotope) of Test Source you

are installing (or replacing with).

• Type in the Test Source Field Strength in uCi (micro-Curies).
• Enter the date on the Test Source.

• Enter the Serial Number from the label on the Test Source.
• Finally, enter your initials, then click on the ‘Done’ button

The information you entered will be entered into the Quality Assessment Event log.