Speed Calibration with Line Down Contact

PROCEDURE
(For weigh scales using line down contact for speed signal)
(If the weigh scale uses encoder or tachometer, see Low Speed Reference and High Speed Calibration
Procedures.)

Using the Ronan local display or HART communicator find the Main Menu;
Scroll down until you reach and highlight Calibration.
Access Calibration by pressing the right arrow key.
Scroll down until you reach and highlight Cal Speed.
Access the Cal Speed by pressing the right arrow key.
Scroll down until you reach and highlight Manual Entry.
Access the Manual Entry by pressing the right arrow key.
Scroll down until you reach and highlight Belt Speed.
Access Belt Speed by pressing the right arrow key.
Use the arrow keys to select and change the display value to the correct speed. Once the displayed value is
correct, press the F4 key (Enter).
The Speed Calibration is complete.