Mechanical Mounting
Review the Configuration Drawing which is
included in the Drawing Chapter of this manual.
Please reference the dimensional drawings located
in the Drawing Chapter of this manual when
installing the equipment.
Consider the following general guidelines when
mounting the sensor and detector:
Avoid internal vessel obstructions such as baffles, agitators,
manways, heater/cooler tubes, etc. which could interfere with
the transmission through the conveyor/screw of the radiation's
"active beam."
The source and detector must be rigidly mounted so they do
not move with respect to each other. Such movement will
destroy the system's calibration and/or its measurement.
Insulation must be used at the point of installation IF:
- the temperature of the vessel at that spot exceeds 131
o
F
(55
o
C), or - the voltage transmission through the vessel could
interfere with the signal transmission from the source to the
detector.