Linearization Menu

The X96S is capable of performing a multi-point linearization of the weight data when required by an application. The
linearization table contains thirty entries, numbered 1 through 30. Each entry consists of a measured value, an actual value, and a
flag that indicates if the entry is used

2

.

The Linearization menu is used to control the linearization mechanism. It contains the following items:

ITEM

FUNCTION

Linearize

Shows the current state of the Linearization mechanism (enabled or disabled) and
allows the user to change the state.

Clear Table

This item invokes a method that clears all entries in the linearization table

Config Linearize

Selecting this item takes the user to the Config Linearize menu

Config Linearize Menu

The Config Linearize menu is used to configure the parameters associated with linearization of the measured data. It contains
the following items:

ITEM

FUNCTION

Table Entry #

Shows, and allows the user to select, an entry in the linearization table

Entry Used

Shows if the entry is used or not.

Measured

Shows, and allows the user to set, the measured value associated with this linearization
table entry. This is the nonlinear value calculated by the X96S when linearization is
disabled.

Actual

Shows, and allows the user to set, the actual value associated with this linearization
table entry. This value is the result of actual level knowledge, and compares to the
Measured value above.

Set Entry

This item invokes a method that sets a table entry

Remove Entry

This item invokes a method that removes a table entry

Scan Time Menu

The

Scan Time

menu is used to configure the rate the input board scans the detector signal and the rate the microprocessor

updates the LCD display and the output signal.

2

Not all of the entries need to be used and the entries do not need to be used in any particular order.