beautypg.com

Fluid nozzle, Ransburg, Vector r series classic applicators - maintenance – Ransburg Vector R Series Classic 79520 R90 Waterborne User Manual

Page 37: Cleaning and inspection, Removal

background image

Vector R Series Classic Applicators - Maintenance

33

Ransburg

AH-06-02.12

Cleaning and Inspection

1. Use a suitable solvent to clean the fluid nozzle.

2. Examine the fluid nozzle for damage to the air

passages and the fluid nozzle tip. Also, exam-

ine the needle seat for damage or wear. If any

damaged or any worn areas are found, the fluid

nozzle must be replaced.

FLUID NOZZLE

(Refer to Figure 24)

Removal

1. Remove the air cap from the applicator (refer to

"Air Cap - Removal" in the "Maintenance" section).

2. Tilt the applicator forward and pull the trigger to

make sure that all fluid in the applicator is drained

out.

3. While holding the applicator assembly with the

trigger pulled back, remove the fluid nozzle with

the open-end of the special multi-purpose wrench.

†

The applicator barrel

MUST be tilted

front down to remove the fluid nozzle. Fail-

ure to do so may allow paint to enter the

air passages, thereby reducing airflow and

damaging the applicator barrel. Applicators

may be flushed in lieu of tilting. However,

they

MUST be either flushed or tilted during

nozzle removal!

C A U T I O N

!

†

If the fluid nozzle is replaced, there is

a good chance that the needle/electrode

assembly will need to be replaced also. A

worn needle/electrode assembly may not

always seat well in a new fluid nozzle.

NOTE

Figure 9: Air Cap, Fluid Nozzle, and Nut

Figure 8: Removal of Fluid Nozzle

†

To keep the needle/electrode from

unscrewing from the needle shaft, the

applicator's trigger should be actuated

to pull the needle/electrode away from

the fluid nozzle.

NOTE

1 - Retaining Ring

2 - O-Ring, PTFE Encapsulated

3 - Air Cap

4 - Fluid Nozzle

5 - Electrode Assembly

6 - Barrel Assembly

1

2

3

4

5

6