Getting Started

Model GFC7001E Carbon Dioxide Analyzer

Teledyne Analytical Instruments

44

3.4.2. PNEUMATIC CONNECTIONS TO GFC 7001E/EM BASIC

CONFIGURATION

NOTE

In order to prevent dust from getting into the gas flow channels of your analyzer, it was shipped with

small plugs inserted into each of the pneumatic fittings on the back panel.

Make sure that all of these dust plugs are removed before attaching

exhaust and supply gas lines.

See Figure 3-2 and Table 3-2 for the location and descriptions of the various pneumatic inlets/outlets referred to
in this section.

See Section 5.6 for information regarding the pneumatic setup of GFC 7001E/EM Analyzers with various optional
calibration valve options in stalled

CAUTION

G

ENERAL

S

AFETY

H

AZARD

Sample and calibration gases should only come into contact with PTFE (Teflon), FEP,

glass, stainless steel or brass.

The exhaust from the analyzer’s internal pump MUST be vented outside the immediate

area or shelter surrounding the instrument.

It is important to conform to all safety requirements regarding exposure to CO.

Figure 3-10:

Pneumatic Connections–Basic Configuration–Using Bottled Span Gas