beautypg.com

Teledyne GFC-7001E - Trace CO Analyzer User Manual

Page 127

background image

Advanced Features

Model GFC7001E Carbon Dioxide Analyzer

Teledyne Analytical Instruments

127

7.3. SETUP MORE DIAG: USING THE DIAGNOSTICS

FUNCTIONS

A series of diagnostic tools is grouped together under the SETUPMOREDIAG menu, as these parameters
are dependent on firmware revision (see Appendix A). These tools can be used in a variety of troubleshooting
and diagnostic procedures and are referred to in many places of the maintenance and trouble-shooting sections
of this manual.

The various operating modes available under the DIAG menu are:

Table 7-5: Diagnostic Mode (DIAG) Functions

DIAG SUBMENU

SUBMENU FUNCTION

Front Panel Mode

Indicator

MANUAL

SECTION

SIGNAL I/O

Allows observation of all digital and analog signals in
the instrument. Allows certain digital signals such as
valves and heaters to be toggled ON and OFF.

DIAG I/O

13.1.3

ANALOG OUTPUT

When entered, the analyzer performs an analog
output step test. This can be used to calibrate a
chart recorder or to test the analog output accuracy.

DIAG AOUT

13.5.7.1

ANALOG I/O

CONFIGURATION

This submenu allows the user to configure the
analyzer’s analog output channels, including
choosing what parameter will be output on each
channel. Instructions that appear here allow
adjustment and calibration of the voltage signals
associated with each output as well as calibration of
the analog to digital converter circuitry on the
motherboard.

DIAG AIO

7.4.1

ELECTRICAL

TEST

When activated, the analyzer performs an electrical
test, which generates a voltage intended to simulate
the measure and reference outputs of the
SYNC/DEMOD board to verify the signal handling
and conditioning of these signals.

DIAG ELEC

9.6.4

13.5.6.2

DARK

CALIBRATION

1

Disconnects the preamp from synchronous
demodulation circuitry on the SYNC/DEMOD PCA to
establish the dark offset values for the measure and
reference channel.

DIAG DARK

CAL

9.6.1

PRESSURE

CALIBRATION

1

Allows the user to calibrate the sample pressure
sensor.

DIAG PCAL

9.6.2

FLOW

CALIBRATION

1

This function is used to calibrate the gas flow output
signals of sample gas and ozone supply.

DIAG FCAL

9.6.3

TEST CHAN

OUTPUT

Selects one of the available test channel signals to
output over the A4 analog output channel.

DIAG TCHN

7.4.6

1

These settings are retained after exiting DIAG mode.