beautypg.com

Teledyne GFC-7001E - Trace CO Analyzer User Manual

Page 275

background image

Troubleshooting & Repair

Model GFC7001E Carbon Dioxide Analyzer

Teledyne Analytical Instruments

275

13.1.4.2. Sync Demodulator Status LED’s

Two LED’s located on the Sync/Demod Board and are there to make it obvious that the GFC Wheel is spinning
and the synchronization signals are present:

Table 13-3: Sync/Demod Board Status Failure Indications

LED

FUNCTION

FAULT STATUS

INDICATED FAILURE(S)

D1

M/R Sensor Status

(Flashes slowly)

LED is stuck

ON or OFF

GFC Wheel is not turning
M/R Sensor on Opto-Pickup Board failed
Sync/Demod Board failed
JP 4 Connector/Wiring faulty
Failed/Faulty +5 VDC Power Supply (PS1)

D2

Segment Sensor

Status

(Flashes quickly)

LED is stuck

ON or OFF

GFC Wheel is not turning
Segment Sensor on Opto-Pickup Board failed
Sync/Demod Board failed
JP 4 Connector/Wiring faulty
Failed/Faulty +5 VDC Power Supply (PS1)

D1 – M/R Sensor Status

D2 – Segment Sensor Status

JP4 Connector to Opto-Pickup

Board

Figure 13-4:

Sync/Demod Board Status LED Locations