Calibration Procedures

Model GFC7001E Carbon Dioxide Analyzer

Teledyne Analytical Instruments

205

9.6. CALIBRATION OF THE GFC 7001E/EM’S ELECTRONIC

SUBSYSTEMS

9.6.1. DARK CALIBRATION TEST

The dark calibration test interrupts the signal path between the IR photo-detector and the remainder of the
sync/demod board circuitry. This allows the instrument to compensate for any voltage levels inherent in the
sync/demod circuitry that might effect the calculation of CO concentration.

Performing this calibration returns two offset voltages, one for CO MEAS and one for CO REF that are
automatically added to the CPU’s calculation routine. The two offset voltages from the last calibration procedure
may be reviewed by the user via the front panel display.

To activate the dark calibration procedure or review the results of a previous calibration, press:

Offset for CO REF signal

SAMPLE

RANGE=50.0 PPM CO= XX.XX

TST> CAL

SETUP

SETUP X.X

SECONDARY SETUP MENU

COMM VARS

DIAG

EXIT

SETUP X.X

PRIMARY SETUP MENU

CFG DAS RNGE PASS CLK

MORE

EXIT

DIAG

SIGNAL I/O

PREV

NEXT

ENTR

EXIT

SETUP X.X

ENTER PASSWORD

8

1

8

ENTR

EXIT

Continue pressing NEXT until ...

DIAG OPTIC

DARK CALIBRATION

PREV NEXT

ENTR

EXIT

DIAG DARK

CO DARK CALIBRATION

VIEW

CAL

EXIT

DIAG DARK

REF DARK OFFSET: 0.0mV

EXIT

Offset for CO MEAS signal

DIAG DARK

MEAS DARK OFFSET: 0.0mV

EXIT

DIAG DARK

DARK CAL 1% COMPLETE

EXIT

DIAG DARK

DARK CALIBRATION ABORTED

EXIT

Calibration runs automatically