Setup  acal: programming and auto cal sequence – Teledyne GFC-7001E - Trace CO Analyzer User Manual

Page 201

background image

Calibration Procedures

Model GFC7001E Carbon Dioxide Analyzer

Teledyne Analytical Instruments

201

9.4.1. SETUP ACAL: PROGRAMMING AND AUTO CAL SEQUENCE

NOTE

If at any time an illegal entry is selected, (for example: Delta Days > 366) the ENTR key will disappear

from the display.

To program the example sequence shown in Table 9-4, press:

CONTINUE NEXT PAGE

With STARTING TIME

SAMPLE RANGE = 50.0 PPM CO=XX.XX


< TST TST > CAL CALZ CZLS

SETUP

SETUP X.X


CFG ACAL DAS RNGE PASS CLK MORE EXIT

SETUP X.X

SEQ 1) DISABLED


NEXT

MODE

EXIT

SETUP X.X

SEQ 2) DISABLED


PREV NEXT

MODE

EXIT

SETUP X.X

MODE: DISABLED


NEXT

ENTR EXIT

SETUP X.X

MODE: ZERO


PREV

NEXT

ENTR EXIT

SETUP X.X

SEQ 2) ZERO–SPAN, 1:00:00


PREV NEXT

MODE

SET

EXIT

SETUP X.X

TIMER ENABLE: ON


SET>

EDIT

EXIT

SETUP X.X

STARTING DATE: 01–JAN–07


EDIT

EXIT

SETUP X.X

STARTING DATE: 01–JAN–02


0

4 SEP

0

8

ENTR

EXIT

Toggle keys to set

Day, Month & Year:

Format : DD-MON-YY

SETUP X.X

MODE: ZERO–SPAN


PREV NEXT

ENTR

EXIT