beautypg.com

Teledyne GFC-7001E - Trace CO Analyzer User Manual

Page 106

background image

Basic Operation

Model GFC7001E Carbon Dioxide Analyzer

Teledyne Analytical Instruments

106

6.6.5. SETUP RNGE DIL: USING THE OPTIONAL DILUTION RATIO

FEATURE

This feature is a optional software utility that allows the user to compensate for any dilution of the sample gas
that may occur before it enters the sample inlet. Typically this occurs in continuous emission monitoring (CEM)
applications where the sampling method used to remove the gas from the stack dilutes it.

Using the dilution ratio option is a 4-step process:

1. Select the appropriate units of measure (see Section 6.6.4).
2. Select the reporting range mode and set the reporting range upper limit (see Section 6.6.3). Make sure

that:
 The upper span limit entered for the reporting range is the maximum expected concentration of the

UNDILUTED gas.

3. Set the dilution factor as a gain (e.g., a value of 20 means 20 parts diluent and 1 part of sample gas):

4. Calibrate the analyzer.

 Make sure that the calibration span gas is either supplied through the same dilution system as the

sample gas or has an appropriately lower actual concentration.

EXAMPLE: If the reporting range limit is set for 100 ppm and the dilution ratio of the sample gas is 20
gain, either:

a span gas with the concentration of 100 ppm can be used if the span gas passes through the same

dilution steps as the sample gas, or;

a 5 ppm span gas must be used if the span gas IS NOT routed through the dilution system.