beautypg.com

8 calibration requirements for independent range, 9 calibration quality – Teledyne 9110AH - Nitrogen oxides analyzer User Manual

Page 98

background image

TELEDYNE Model 9110AH NO

X

Analyzer Operator Manual, 01620, Rev. F

7-10

Table 7-8: Calibration Requirements for AutoRange or Remote Range

No matter which calibration method you choose from this chapter, all of the following
parameters must be set to properly calibrate the instrument on AutoRange or Remote range.

Parameter

Low Range

High Range

Range

Low Range is selected by pressing
CONC-LO on the front panel.

High Range is selected by pressing
CONC-HI on the front panel

Slope

NO

x

and NO slopes for low range

are set by span calibration

NO

x

and NO slopes for high range

are set by span calibration

Offset

NO

x

and NO offsets for low range

are set by zero calibration

NO

x

and NO offsets for high range

are set by zero calibration

Converter Efficiency

CE can be set by one of the
procedures in Section 7.10.
Converter efficiency is optional.

CE can be set by one of the
procedures in Section 7.10.
Converter efficiency is optional.

7.8 Calibration Requirements for Independent Range

In Independent Range mode there is one set of calibration parameters. That is, NO

x

and NO

slopes and NO

x

and NO offsets, and one converter efficiency parameter. The calibration of the

M9110AH using independent range is the same as calibration in single range. To achieve
independent ranges, the computer scales the 3 outputs using the ranges keyed in by the
operator.

7.9 Calibration Quality

After calibration is complete, it is very important to check the QUALITY of the calibration. The
calibration of the M9110AH involves balancing several sections of electronics and software to
achieve optimum accuracy, noise, linearity and dynamic range.

The slope and offset parameters are similar to the span and zero pots on an analog instrument.
Just as in the analog instrument, if the slope or offset get outside of a certain range, the
instrument will not perform as well.

The slope value will be slightly different on the NO and NO

x

channels. This is due to slight

differences in pneumatic resistance in each pathway. If the slopes are significantly different,
there is a calibration error or a cross port leak in the switching valve. If there is a sudden
change in slopes after a calibration, that usually indicates a change in reaction cell pressure.
Check the Troubleshooting Section 9.