TELEDYNE Model 9110AH NO

X

Analyzer Operator Manual, 01620, Rev. F

9-39

9.3.3.1 ADC/DAC Calibration Procedure

Due to the stability of modern electronics, this procedure should only need to be performed if a
major sub-assembly is exchanged or when the display voltage does not match the input voltage
or current to the V/F card. After completion, a Factory Calibration Procedure should be
performed, see Section 9.1.6.

Before the actual calibration is performed, switches on the V/F card must be correctly set and
jumpers set on the motherboard. Jumper and switch setting changes must be performed
with the instrument power OFF.

Motherboard Jumpers

The motherboard contains 4 pairs of jumpers JP1 - JP8, one pair for each analog output
channel. Each channel can be configured for either voltage or current output. Use Table 9-7 to
configure the jumpers.

Table 9-7: Motherboard Jumper Settings

Analog output

Terminal Pair
Rear panel

Jumper Pair

Jumper Setting for
Voltage Mode

Jumper Setting for
Current Mode

DAC 0 - NO

x

3-4

JP3 - JP4

B-C

A-B

DAC 1 - NO

5-6

JP1 - JP2

B-C

A-B

DAC 2 - NO2

1-2

JP5 - JP6

B-C

A-B

DAC 3 - TEST

7-8

JP7 - JP8

B-C

A-B

V/F Board Switch Settings

There are 2 different types of current outputs, Non-Isolated (std equipment) and Isolated. Each
requires a different switch setting shown below. If you are operating the instrument in voltage
output mode, the switches should be set to the desired voltage range.

Table 9-1: V/F Board Switch Settings

DAC #

Sw 1

Sw 2

Sw 3

Sw 4

Sw 5

Sw 6

Sw 7

0

ON*

OFF*

OFF*

1

ON*

OFF*

OFF*

2

ON*

OFF*

OFF*

3

ON*

OFF*

OFF*

*Required settings