TELEDYNE Model 9110AH NO
X
Analyzer Operator Manual, 01620, Rev. F
5-30
Table 5-12: RS-232 Command Summary
Commands
Definition
?
Print help screen
T SET ALL
Enable display of all test variables during T LIST
T SET name
Display only NAME during T LIST
T LIST [ALL]
Print all test variables enabled with T SET or ALL warnings
T name
Print single test, "name" from Table 5-15
T CLEAR ALL
Disable T LIST, use with T SET name
W SET ALL
Enable display of all warnings during W LIST
W LIST [ALL]
Print warnings enabled with W SET or ALL warnings
W name
Print individual "name" warning from Table 5-16
W CLEAR ALL
Disable W LIST, use with W SET
C LIST
Print calibration commands
C command
Execute calibration "command" from Table 5-18
D LIST
Prints all I/O signal values
D name
Prints single I/O signal value/state
D name=value
Sets variable to new "value"
D LIST NAMES
Lists diagnostic test names
D ENTER name
Enters and starts 'name' diagnostic test
D EXIT
Exits diagnostic mode
D SYS_RESET
Resets analyzer(same as power-on)
D RAM_RESET
System reset, plus erases RAM. Initializes DAS, NO, NO
x
, NO
2
conc
readings, calib not affected.
D EE_RESET
System reset, plus erases EEPROM (RAM_RESET actions + setup
variables, calibration to default values.
V LIST
Print all variable names
V name
Print individual "name" variable
V name=value
Sets variable to new "value"
V CONFIG
Print analyzer configuration
R count
Print last "count" DAS averages