beautypg.com

3 specifications, warranty, 1 specifications – Teledyne 9110AH - Nitrogen oxides analyzer User Manual

Page 29

background image

TELEDYNE Model 9110AH NO

X

Analyzer Operator Manual, 01620, Rev. F

3-1

3 SPECIFICATIONS, WARRANTY

3.1 Specifications

Operating Modes

NO/

NO

x

switching mode, NO only mode,

NO

x

only mode

Ranges

In 1 ppm increments from 5 ppm to 5,000 ppm

Single range, independent ranges or autoranging

Noise at zero

0.02 ppm RMS

Noise at span

<0.2% of reading RMS above 20 ppm

Detection Limit(Note 1)

0.04 ppm RMS

Zero Drift (Note 2)

<0.2% full scale/24 hours

Zero Drift (Note 2)

<0.4% full scale/7 days

Span Drift (Note 2)

<1% FS/24 hours

Lag Time

Switching Mode

20 sec (Note 3)

NO

x

mode

4 sec (Note 3)

Response Time

Switching Mode

95% in < 40 sec (Note 3)

NO

x

mode

95% in < 10 sec (Note 3)

Sample Flow Rate

290 10 cc/min (Including bypass)

Linearity

1% of full scale

Precision

0.5% of reading

Temperature Range

5-40

o

C

Humidity

0-95% RH non-condensing

Temp Coefficient

< 0.1% per

o

C

Voltage Coefficient

< 0.1% per V

Dimensions HxWxD

7"x17"x23.6" (18 cm x 43 cm x 61 cm)

Weight, Analyzer

43 lbs (20 kg)

Weight, Pump Pack

16 lbs (7 kg)

Power, Analyzer

100 V~ 50/60 Hz, 120 V~ 60 Hz, 220 V~ 50 Hz, 240 V~ 50 Hz, 200 watts

Power, Analyzer

4

230 V~ 50 Hz, 2.5A

Power, Ext Pump

110 V~ 60 Hz, 220 V~ 50 Hz, 240 V~ 50 Hz, 295 watts

Power, Ext Pump

4

230 V~ 50 Hz, 2.5A

Environmental

Installation Category (Over-voltage Category) II

Pollution Degree 2

Analog Resolution

1 part in 2048 of selected voltage or current range

Recorder Output

0-100 mV, 0-1, 5, 10v, bipolar

Current Loop Option

4-20ma isolated

Status

12 Status Outputs from opto-isolator

Measurement Units

ppm, mg/m

3

1. Defined as twice the zero noise level.
2. At constant temperature and voltage.
3. Lag & response times longer for external converter option.
4. Electrical ratings for CE Mark compliance.