TELEDYNE Model 9110AH NO

X

Analyzer Operator Manual, 01620, Rev. F

5-40

Where:

name

= name of the variable

value

= current value of variable

warning_lo

= lower limit warning (displayed if applicable)

warning_hi

= upper limit warning (displayed if applicable)

data_lo

= lower limit of allowable values

data_hi

= upper limit of allowable values

Variables can be changed. Before changing the settings on any variables, please make sure you
understand the consequences of the change. We recommend you call the factory before
changing the settings on any variables. The general format for changing the settings on a
variable is:

V LIST name[=value [warn_lo [warn_hi]]]

For example to change the warning limits on the box temperature type:

V BOX_SET 30 10 50

and the CPU should respond with:

V DDD:HH:MM IIII BOX_SET=30 10 50(0-60)

The CONFIG command lists the software configuration.

For example:

V CONFIG

list software configuration

The format of this listing is shown in the example below.

CONFIG[0]

= Revision H.7

CONFIG[1]

= NO

x

Analyzer

CONFIG[2]

= SBC40 CPU