TELEDYNE Model 9110AH NO

X

Analyzer Operator Manual, 01620, Rev. F

2-13

Table 2-1: Final Test and Calibration Values

TEST Values

Observed
Value

Units

Nominal Range

Reference Section

RANGE

ppm

5-5000

5.3.4

NOISE

ppm

0.0 - 0.2

9.1.1, Table 9-1, 9.2.5

SAMP FLW

cc/min

300 50 (Default)
550 50 (Optional)

9.3.7, Table 9-1

OZONE FL

cc/min

250 15

9.3.6

PMT

mV

0-5000

9.3.8

AUTOZERO

mV

-10 to +50

4.1

HVPS

V

400 - 700 constant

9.3.8.5

DCPS

mV

2500 200

9.3.5

RCELL TEMP

o

C

50 2

9.3.8.2

BLOCK TEMP

o

C

50 2

9.3.4.1

BOX TEMP

o

C

8-48

9.3.4.1

PMT TEMP

o

C

7 1

9.3.8.4

CONV TEMP

o

C

700 10 (Std)
315 5 (Moly)

9.3.4.1

RCEL PRES

IN-Hg-A

2 - 10 constant

9.3.7

SAMP PRES

IN-Hg-A

25 - 30 constant

9.3.7

Electric Test &Optic Test

Electric Test

PMT Volts

mV

2000 200

9.1.3.2

NO Conc

ppm

250 25

9.1.3.2

NO

x

Conc

ppm

250 25

9.1.3.2

OPTIC TEST

PMT Volts

mV

100 20

9.1.3.3

NO Conc

ppm

12.5 2

9.1.3.3

NO

x

Conc

ppm

12.5 2

9.1.3.3

(table continued)