Teledyne 9110AH - Nitrogen oxides analyzer User Manual

Page 80

background image

TELEDYNE Model 9110AH NO

X

Analyzer Operator Manual, 01620, Rev. F

5-38

To print records from a DAS data channel enter:

D REPORT “name” RECORDS=nnn COMPACT|VERBOSE

Examples of reports are:

D REPORT “CONC” RECORDS=35 VERBOSE

D REPORT “CALDAT” RECORDS=10

D REPORT “PNUMTC” RECORDS=155 VERBOSE

Automatic RS-232 reporting can be independently enabled and disabled for each Data Channel.
For all default data channels, automatic reporting is initially set to “OFF.” If this property is
turned on, the Data Channel will issue a report with a time and date stamp to the RS-232 port
every time a data point is logged. The report format is shown below:

D 94:08:00 0200 CONC : AVG NXCNC1 = 1234.5 PPB

D 94:08:00 0200 CONC : AVG NOCNC1 = 1234.5 PPB

D 94:08:00 0200 CONC : AVG N2CNC1 = 1234.5 PPB

One CONC report consists of:

D

= Type of report (Diagnostic)

94:08:00

= Time and Date stamp (Julian day, Hr, Min)

0200

= Instrument ID number

CONC

= Data Channel name

CONC = concentration data

PNUMTC = pneumatic parameters

CALDAT = calibration parameters

AVG

= Type of data

AVG = average reading

INST = instantaneous reading

NXCNC1 = 1234.5 PPB

= Name of the parameter

NX = NOx,

NO = NO

N2 = NO2.