8 variables commands and messages – Teledyne 9110AH - Nitrogen oxides analyzer User Manual
Page 81
![background image](/manuals/345703/81/background.png)
TELEDYNE Model 9110AH NO
X
Analyzer Operator Manual, 01620, Rev. F
5-39
All of the default Data Channels sample more than one parameter, for these channels, each
parameter is printed on a separate line.
There is also a compact format. If this attribute is enabled, all 3 concentration parameters are
printed on one line as shown below:
D 94:08:00 0200 CONC : 120.0 100.0 20.0
The parameters are in the order of NO
x
, NO, and NO
2
.
To change any of the attributes of a particular data channel, the channel attributes are edited
from the front panel. The following table uses the example of Automatic Reporting. Other
attributes can be edited in a similar fashion
5.7.8 VARIABLES Commands and Messages
The M9110AH operational modes are listed above. To list the analyzer's current mode type:
V MODE
Lists M9110AH current operational mode
See Table 5-7 for list of operational modes and meanings.
Model M9110AH Internal variables
The M9110AH has a number of internal setup variables. Essentially all of these are set at time of
manufacture and should not need to be changed in the field. A list of user accessible variables
is shown in Table 9-5.
A list of variables and their settings can be requested over the RS-232 port by:
V LIST
Lists internal variables and values
The output from this command is quite long and will not be shown here. The general format of
the output is:
name = value warning_lo warning_hi