Teledyne 9110AH - Nitrogen oxides analyzer User Manual
Page 134
![background image](/manuals/345703/134/background.png)
TELEDYNE Model 9110AH NO
X
Analyzer Operator Manual, 01620, Rev. F
9-18
Table 9-5: Model 9110AH Variables
No.
Name
Units
Default
Value
Value
Range
Description
0
DAS_HOLDOFF
min
15
0.5 - 20
Time, in minutes, that data is not
stored in the DAS after
calibration or power-up.
1
MEASURE MODE
text
NONOX
NO, NO
x
,
NONOX
NO = only mode
NO
x
= only mode
NO/NO
x
= switching mode
2
TPC_ENABLE
Logic
ON
ON-OFF
Temp/Pres compensation enable.
3
DYN_ZERO
Logic
OFF
ON-OFF
Calibrates instrument during
AutoCal Sequence or external
contact closure or RS-232
command
4
DYN_SPAN
Logic
OFF
ON-OFF
Calibrates instrument during
AutoCal Sequence or external
contact closure or RS-232
command
5
CONC_LIM_ENA
LOGIC
NO
NO/YES
Enable concentration limit alarms
6
NOx_CONC_LIM
PPM
0
5000
NO
x
concentration limit alarm
7
NO_CONC_LIM
PPM
0
5000
NO concentration limit alarm
8
NO2_CONC_LIM
PPM
0
5000
NO
2
concentration limit alarm
9
SFLOW_SET
cc/min.
290
0-1000
Nominal sample flow rate
10
OFLOW_SET
cc/min.
250
0-500
Nominal ozone flow rate
11
RS232_MODE
Bit
Field
8
0-99999
RS-232 port 1 mode flags. Add
values to combine flags.
1 = quiet mode
2 = computer mode
4 = enable security
8 = enable TELEDYNE protocol
16 = enable alternate protocol
32 = enable multidrop
64 = enable modem
128 = ignore RS-232 line errors
4096 = enable command prompt
12
CLOCK_ADJ
Sec.
0
+-60
Real-time clock speed adjustment