Teledyne 9110AH - Nitrogen oxides analyzer User Manual

Page 134

background image

TELEDYNE Model 9110AH NO

X

Analyzer Operator Manual, 01620, Rev. F

9-18

Table 9-5: Model 9110AH Variables

No.

Name

Units

Default
Value

Value
Range

Description

0

DAS_HOLDOFF

min

15

0.5 - 20

Time, in minutes, that data is not
stored in the DAS after
calibration or power-up.

1

MEASURE MODE

text

NONOX

NO, NO

x

,

NONOX

NO = only mode
NO

x

= only mode

NO/NO

x

= switching mode

2

TPC_ENABLE

Logic

ON

ON-OFF

Temp/Pres compensation enable.

3

DYN_ZERO

Logic

OFF

ON-OFF

Calibrates instrument during
AutoCal Sequence or external
contact closure or RS-232
command

4

DYN_SPAN

Logic

OFF

ON-OFF

Calibrates instrument during
AutoCal Sequence or external
contact closure or RS-232
command

5

CONC_LIM_ENA

LOGIC

NO

NO/YES

Enable concentration limit alarms

6

NOx_CONC_LIM

PPM

0

5000

NO

x

concentration limit alarm

7

NO_CONC_LIM

PPM

0

5000

NO concentration limit alarm

8

NO2_CONC_LIM

PPM

0

5000

NO

2

concentration limit alarm

9

SFLOW_SET

cc/min.

290

0-1000

Nominal sample flow rate

10

OFLOW_SET

cc/min.

250

0-500

Nominal ozone flow rate

11

RS232_MODE

Bit
Field

8

0-99999

RS-232 port 1 mode flags. Add
values to combine flags.
1 = quiet mode
2 = computer mode
4 = enable security
8 = enable TELEDYNE protocol
16 = enable alternate protocol
32 = enable multidrop
64 = enable modem
128 = ignore RS-232 line errors
4096 = enable command prompt

12

CLOCK_ADJ

Sec.

0

+-60

Real-time clock speed adjustment