beautypg.com

Teledyne 9110AH - Nitrogen oxides analyzer User Manual

Page 156

background image

TELEDYNE Model 9110AH NO

X

Analyzer Operator Manual, 01620, Rev. F

9-40

10 V output or non-isolated current loop

Switch

State

Comment

3

ON

Use for non-isolated current loop or 10 V output

4

OFF

Use for non-isolated current loop or 10 V output

5

OFF

Use for non-isolated current loop or 10 V output

6

OFF

Use for non-isolated current loop or 10 V output

5 V output or isolated current loop

Switch

State

Comment

3

OFF

Use for isolated current loop or 5 V output

4

ON

Use for isolated current loop or 5 V output

5

OFF

Use for isolated current loop or 5 V output

6

OFF

Use for isolated current loop or 5 V output

1 V output

Switch

State

Comment

3

OFF

Use for 1 V output

4

OFF

Use for 1 V output

5

ON

Use for 1 V output

6

OFF

Use for 1 V output

100 mV output

Switch

State

Comment

3

OFF

Use for 100 mV output

4

OFF

Use for 100 mV output

5

OFF

Use for 100 mV output

6

ON

Use for 100 mV output