Agilent Technologies E8267D PSG User Manual
Page 64

50
Chapter 2
Basic Operation
Configuring the RF Output
Using Basic Ramp Sweep Functions
This procedure demonstrates the following tasks (each task builds on the previous task):
•
“Configuring a Frequency Sweep” on page 50
•
•
“Adjusting Sweep Time” on page 54
•
“Using Alternate Sweep” on page 55
•
“Configuring an Amplitude Sweep” on page 56
Configuring a Frequency Sweep
1.
Set up the equipment as shown in
NOTE
The PSG signal generator is not compatible with the GPIB system interface of an 8757A,
8757C, or 8757E. For these older scalar network analyzers, do not connect the GPIB cable in
. These scalar network analyzers provide only a subset of 8757D functionality. See
the Data Sheet for details. Use the 8757A/C/E documentation instead of this procedure.
Figure 2-3
Equipment Setup
2.
Turn on both the 8757D and the PSG.
3.
On the 8757D, press
System
>
More
>
Sweep Mode
and verify that the
SYSINTF
softkey is set to ON.
This ensures that the system interface mode is activated on the 8757D. The system interface