Chapter 3

101

Basic Digital Operation

Using Waveform Markers

Accessing Marker Utilities

Use the following procedure to display the marker parameters. This procedure uses the Dual ARB
player, but you can access the marker utilities through the

ARB Setup

softkey in all ARB formats.

1.

Select the ARB waveform player:

press

Mode

>

Dual ARB

2.

Press

ARB Setup

>

Marker Utilities

.

Sample range begins on first point of signal

Negative range set between signal and

off time

Example of Incorrect Use

Waveform: 1022 points

Marker range: 110–1022

Marker polarity: Negative

This figure shows that a negative polarity marker goes low during
the marker on points; the marker signal goes high during the off
points. The ALC samples the waveform during the off marker
points.

Marker On

Marker On

Sampling both on and off time sets the modulator circuitry
incorrectly for higher signal levels. Note the increased amplitude
at the beginning of the pulse.

Marker

Off

Marker On

Marker

Marker

Off

On