beautypg.com

Operating the n5102a module in output mode – Agilent Technologies E8267D PSG User Manual

Page 269

background image

Chapter 12

255

Peripheral Devices

N5102A Digital Signal Interface Module

The

Filter

softkey accesses a menu that enables you set the desired filtering parameters.

Operating the N5102A Module in Output Mode

This section shows how to set the parameters for the N5102A Option 003 module in output mode
using the front panel keys. Each procedure contains a figure that shows the softkey menu structure
for the interface module function being performed.

Setting up the Signal Generator Baseband Data

The digital signal interface module receives data from a baseband source and outputs a digital IQ or
digital IF signal relative to the selected logic type. Because the PSG provides the baseband data, the
first procedure in operating the interface module is configuring the PSG using one of the real- time or
ARB modulation formats, or playing back a stored file using the Dual ARB player. For information on
setting up real- time or ARB waveforms, or to learn about using the Dual ARB player, refer to the
appropriate chapter in this guide.

1.

Preset the signal generator.

2.

Select the modulation format (TDMA, Custom, and so forth) and set the desired parameters.

3.

Turn- on the modulation format.

Accessing the N5102A Module User Interface

Press

Aux Fctn

>

N5102A Interface

.

This accesses the UI (first- level softkey menu shown in

Figure 12- 9

) that is used to configure the

digital signal interface module. Notice the graphic in the PSG display showing a setup where the
N5102A module is generating its own internal clock signal. This graphic changes to reflect the
current clock source selection.

Table 12-7 Maximum Sample Rate for Selected Filter

Filter

Maximum Rate

Gaussian

Nyquist

Root Nyquist

Rectangle

Edge

UN3/4 GSM Gaussian

IS- 95

IS 95 w/EQ





50 MHz

IS- 95 Mod

IS- 95 Mod w/EQ

25 MHz

APCO 25 C4FM

12.5 MHz