Optimizing Performance
Using the Option 521 Detector Calibration (Option 521)

Recalling and Applying a User Flatness Correction Array

Before performing the steps in this section, complete the section

“Creating a User Flatness Correction

Array with a mm–Wave Source Module” on page 140

.

1.

Press

Preset

.

2.

Press

Amplitude

>

More (1 of 2)

>

User Flatness

>

Configure Cal Array

>

More (1 of 2)

>

Preset List

>

Confirm Preset

.

3.

Press

More (2 of 2)

>

Load/Store

.

4.

Ensure that the file FLATCAL2 is highlighted.

5.

Press

Load From Selected File

>

Confirm Load From File

.

This populates the user flatness correction array with the data contained in the file FLATCAL2.
The user flatness correction array title displays User Flatness: FLATCAL2.

6.

Press

Return

>

Flatness Off On

.

This activates flatness correction using the data contained in the file FLATCAL2.

Using the Option 521 Detector Calibration (Option 521)

CAUTION

Using the Option 521 Detector Calibration softkey changes the factory calibration. Avoid
using this feature unless following a service center’s recommendations.

This calibration should be done whenever the instrument temperature changes beyond
specified parameters or and when greater amplitude flatness is required around the
500 MHz frequency transition. Refer to the Data Sheet.

NOTE

If the Option 521 Detector Calibration is inadvertently used, the PSG has the ability to recall
the factory amplitude flatness calibration. Refer to

“Restoring the Factory Flatness

Calibration”

.

The purpose of this calibration is to improve the 500 MHz frequency transition to compensate for
changes in temperature. The 500 MHz frequency transition can degrade as the temperature changes.
By using the detector calibration, when the instrument temperature change occurs, the PSG’s flatness
performance can be optimized. Refer to the Data Sheet.

Running the Option 521 Detector Calibration

Following the following steps to run the Detector calibration:

Press

Amplitude > More > More > Option 521 Detector Calibration > Execute

Restoring the Factory Flatness Calibration

The instrument’s original factory calibration settings can be restored by the following steps:

Press

Utility > Power On/Preset > Restore System Defaults