Chapter 4

143

Optimizing Performance

Creating and Applying User Flatness Correction

Figure 4-6

User Flatness with mm–Wave Source Module for Signal Generators with Options 1EA, 1EU, or
521

CAUTION

Option 521 signal generators can damage MM source modules. Consult the MM source
module’s operating manual for input damage levels.

NOTE

To ensure adequate RF amplitude at the mm–wave source module RF input when using
Options 1EA, 1EU, or 521 signal generators, maximum amplitude loss through the adapters
and cables connected between the signal generator’s RF output and the mm–wave source
module’s RF input should be less than 1.5 dB.

Configure the Signal Generator

1.

Turn on the signal generator’s line power. At power–up, the signal generator automatically does
the following:

senses the mm–wave source module

switches the signal generator’s leveling mode to external/source module

sets the mm–wave source module frequency and amplitude to the source module’s preset
values

displays the RF output frequency and amplitude available at the mm–wave source module
output

The MMMOD indicator in the FREQUENCY area and the MM indicator in the AMPLITUDE area of the
signal generator’s display indicate that the mm–wave source module is active