beautypg.com

Agilent Technologies E8267D PSG User Manual

Page 260

background image

246

Chapter 12

Peripheral Devices
N5102A Digital Signal Interface Module

Clock Timing for Parallel Data

Some components require multiple clocks during a single sample period. (A sample period consists of
an I and Q sample). For parallel data transmissions, you can select one, two, or four clocks per
sample. For clocks per sample greater than one, the I and Q samples are held constant to
accommodate the additional clock periods. This reduces the sample rate relative to the clock rate by
a factor equal to the clocks per sample selection. For example, when four is selected, the sample rate
is reduced by a factor of four (sample rate to clock rate ratio).

Figure 12- 4

demonstrates the clock

timing for each clocks per sample selection. For input mode, the clocks per sample setting is always
one.

Externally Supplied Clock

NOTE: Use only one of the two signal generator frequency reference inputs.

This manual is related to the following products: