Agilent Technologies E8267D PSG User Manual
Page 260

246
Chapter 12
Peripheral Devices
N5102A Digital Signal Interface Module
Clock Timing for Parallel Data
Some components require multiple clocks during a single sample period. (A sample period consists of
an I and Q sample). For parallel data transmissions, you can select one, two, or four clocks per
sample. For clocks per sample greater than one, the I and Q samples are held constant to
accommodate the additional clock periods. This reduces the sample rate relative to the clock rate by
a factor equal to the clocks per sample selection. For example, when four is selected, the sample rate
is reduced by a factor of four (sample rate to clock rate ratio).
demonstrates the clock
timing for each clocks per sample selection. For input mode, the clocks per sample setting is always
one.
Externally Supplied Clock
NOTE: Use only one of the two signal generator frequency reference inputs.