289

Troubleshooting

Error Messages

CAUTION

Carefully read the entire message! It may list additional risks with this procedure.

3.

Release the

Preset

key.

4.

To continue with the sequence, press

Continue

(to abort with no lost files, press

Abort

).

5.

When the sequence concludes, do the following:

a.

Cycle power.

Cycling power restores all previously installed options. Because calibration files are restored
from EEPROM, you should see several error messages.

b.

Perform the DCFM/DC

M calibration.

Refer to the

DCFM/DC

M Cal

softkey description in the Agilent PSG Signal Generators Key

Reference.

c.

Agilent Technologies is interested in the circumstances that made it necessary for you to
initiate this procedure. Please contact us at the telephone number listed at
http://www.agilent.com/find/assist. We would like to help you eliminate any repeat
occurrences.

Error Messages

If an error condition occurs in the signal generator, it is reported to both the front panel display
error queue and the SCPI (remote interface) error queue. These two queues are viewed and managed
separately; for information on the SCPI error queue, refer to the Agilent Signal Generators
Programming Guide
.

NOTE

When there is an unviewed message in the front panel error queue, the ERR annunciator
appears on the signal generator’s display.

Characteristic

Front Panel Display Error Queue

Capacity (#errors)

30

Overflow Handling

Circular (rotating).

Drops oldest error as new error comes in.

Viewing Entries

Press:

Utility

>

Error Info

>

View Next

(or

Previous

)

Error Page

Clearing the Queue

Press:

Utility

>

Error Info

>

Clear Error Queue(s)

Unresolved Errors

a

a.Errors that must be resolved. For example, unlock.

Re–reported after queue is cleared.

No Errors

When the queue is empty (every error in the queue has been read, or the queue is cleared), the
following message appears in the queue:

0 No Error Message(s) in Queue